Standard delivery Environmental test chambers
On this page, you can find all the essential technical data on the Memmert environmental chamber CTC/TTC for climate testing and temperature testing.
Temperature
Setting accuracy temperature | up to 99.9 °C: 0.1 / from 100 °C: 0.5 |
---|---|
Temperature range | without humidity: from -42°C up to +190°C |
Temperature range | with humidity from +10°C up to +95°C |
Temperature distribution (spatial) | +/- 0.5 up to 2K |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Display | resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED) |
Humidity
Humidification | humidification by hot steam generator |
---|---|
Humidity control | active humidifying and de-humidifying adjustable from 10-98 % rh with digital display of relative humidity – resolution of display 0.5 %, setting accuracy 1 % |
Humidification | By hot steam generator. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7) from 2 x 10 l tank on telescopic slide integrated in chamber, with automatic change-over. |
Control technology
Controller | Electronic microprocessor temperature controller with auto-diagnostic system |
---|---|
Set-up Display | digital display of all set parameters, such as temperature, weekdays, time, humidity, fan speed, programme status and set-up values – language to be chosen via set-up |
Timer | integrated timer for tempering profiles of up to 40 ramps each, each segment adjustable from 1 min. to 999 hrs. |
Calibration | three freely selectable temperature values |
Ventilation
Fan | high-performance air fan in working chamber |
---|---|
Fan control | depending on operation status automatic adaption of fan speed resp. manual adjustment from 10 to 100 % |
Communication
Interface USB | USB-interface incl. Memmert software “Celsius” for programming and documentation |
---|---|
Interface Printer | parallel printer interface (incl. real time clock with date function) for all PCL3-compatible ink jet printers for GLP-conforming documentation |
Documentation | integrated ring memory as data logger for GLP-conforming long-term documentation of all relevant parameters – 1024 kB |
Documentation | programme stored in case of power failure |
Programming | chip-card control incl. 1 MEMoryCard XL with 32 kB storage capacity (max. 40 ramps) |
Safety
Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
---|---|
Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, adjustable at controller |
AutoSAFETY | additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
Alarm | with visual and acoustic alarm in case of over-/under temperature and underhumidity, open door and empty water tank |
Heating concept
Ring heater | high-performance ring heaters with optimised air circulation |
---|---|
Door | door heating to avoid condensation |
Cooling | twin compressor; refrigerant R449A |
Standard equipment
Entry port | Entry port right, 80 mm, with stopper |
---|---|
Internals | 1 stainless steel grid(s), electropolished |
Works calibration certificate | standard values +10 °C and +37 °C, 60 % rh at +30 °C |
Door | fully insulated stainless steel door with double-locking and 4-point adjustment, heated |
Interior | easy-to-clean stainless steel interior, materil 1.4301 (ASTM 304), hermetically welded |
Stainless steel interior
Volume | 256 l |
---|---|
Dimensions | w(A) x h(B) x d(C): 640 x 670 x 597 mm |
Max. number of internals | 6 |
Max. loading per internal | 25 kg |
Textured stainless steel casing
Dimensions | w(D) x h(E) x d(F): 898 x 1730 x 1100 mm (d +50mm door handle) |
---|---|
Installation | on lockable castors |
Housing | rear zinc-plated steel |
Electrical data
Electrical load | 400 V, 3 phases, 50/60 Hz / approx. 7000 W |
---|
Ambient conditions
Set Up | The distance between the test chamber and the walls/ceiling must be at least 25 cm. |
---|---|
Ambient temperature | 16 ºC to 28 ºC |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Pollution degree | 2 |
Packing/shipping data
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | w x h x d: 1020 x 1910 x 1310 mm |
Net weight | approx. 337 kg |
Gross weight carton | approx. 463 kg |